Software error in Siemens cardiac imaging systems causes inaccurate flow measurements
A software defect in syngo.via MR Cardiac Analysis affects the evaluation of 2D cardiac flow data. The algorithm fails to include the final cardiac phase in calculations, leading to systematic underestimation of flow quantification parameters like forward volume. This issue specifically impacts beat sensor-triggered data where flow peaks occur during the final cardiac phases.
If you use these imaging systems, be aware that cardiac flow measurements may be underestimated and require manual verification of data.